Bowie-Dick Test Failures: What They Indicate—and What They Do Not

Why the Bowie-Dick test matters
In a dynamic-air-removal steam sterilizer, residual air can interfere with steam contact at surfaces that must be sterilized. A Bowie-Dick-type test is designed to challenge the sterilizer’s ability to remove air and allow steam to penetrate a defined test load.[1][2]
For vacuum-assisted steam sterilizers, CDC guidance describes performing an air-removal test each day the sterilizer is used, before the first processed load.[1] The test is an important readiness check, but it is not a substitute for reviewing cycle records or applying the other monitoring and quality controls required by the facility’s procedures.
What a Bowie-Dick result evaluates
The test uses a defined test load together with a chemical indicator system. The indicator is intended to show whether steam penetrated the test load uniformly. A nonuniform color change may indicate inadequate air removal or another condition that interfered with steam penetration.[2][3]
Possible contributors to a failed or abnormal result include:
- inadequate air removal during the cycle;
- air leakage or air re-entrainment;
- non-condensable gases in the steam supply;
- a problem with the test pack or indicator system;
- incorrect test placement, cycle selection, or handling; or
- a sterilizer or utility-system malfunction.
The result can signal a steam-penetration problem, but it generally does not identify the specific cause. ISO 11140-4:2020, Type 2 indicator systems as an alternative to the Bowie and Dick-type test, is the third edition of the standard and cancels and replaces ISO 11140-4:2007.[2] It addresses the performance of the indicator system together with its specified test load. It should not be treated as a general diagnostic procedure for every possible sterilizer fault.
What the test does not prove
It is not a complete sterilization test
A passing Bowie-Dick result does not demonstrate that every load has achieved sterilization. The test uses an otherwise empty chamber and a defined challenge. It does not verify the performance of every load configuration, package, device, or lumen.[2][3]
FDA guidance describes air-removal indicators as tools used to evaluate air removal and steam penetration in dynamic-air-removal sterilizers. It also discusses the role of different monitoring methods rather than treating one indicator result as a complete assessment of the sterilization process.[3]
It does not replace other monitoring
A passing result does not replace mechanical monitoring, review of the sterilizer printout or electronic record, chemical indicators, or biological indicators when they are specified by the facility’s procedure or applicable requirements. These controls address different questions, including whether cycle conditions were recorded, whether an indicator responded to specified process conditions, and whether the process was challenged in an appropriate way.[1][3]
A practical response to a failed test
The workflow below is a facility-process framework for organizing a response. It is not a universal regulatory recall algorithm. Local requirements, the sterilizer manufacturer’s instructions, test-system instructions, and the facility’s quality system determine the actions that apply.
1. Stop routine use of the sterilizer
Do not treat a failed Bowie-Dick result as only a paperwork exception. CDC guidance states that a dynamic-air-removal sterilizer should not be used until it has been inspected and passes the air-removal test.[1] Place the sterilizer in a clearly identified hold or out-of-service status in accordance with the facility’s procedure.
2. Preserve the available evidence
As a facility-process recommendation, retain the failed test or test device when applicable, the cycle record, sterilizer identification, date and time, operator information, and any relevant alarm or maintenance message. A photograph of the indicator may help preserve the observed pattern, but it should supplement rather than replace the original record. The specific records retained should follow the facility’s quality system and the test-system instructions.
3. Check for test-use errors without dismissing the result
Review whether the test was:
- run in an empty chamber;
- performed in the correct sterilizer and cycle;
- positioned as specified by the test-system instructions and facility procedure;
- performed before the first processed load, consistent with the applicable sterilizer procedure and CDC guidance;
- within its expiry date and stored as specified; and
- interpreted within the manufacturer’s stated reading window.
These checks help determine whether the test was valid. They should not be used automatically to explain away a clearly failed result, particularly when the failure recurs or is accompanied by alarms or other evidence of abnormal operation.
4. Escalate the issue through the facility’s technical process
Repeated failures, unusual indicator patterns, alarms, vacuum abnormalities, or suspected steam-supply problems should be referred to qualified technical personnel under the facility’s maintenance and quality procedures. Sterile processing staff should not make unauthorized adjustments to vacuum systems, valves, door seals, steam supplies, or control settings.
Depending on the facility’s procedures and the sterilizer manufacturer’s instructions, the technical review may consider preventive-maintenance status, leak-testing information, chamber and drain condition, vacuum-system performance, steam quality, and recent repairs. These are examples of investigation topics, not a universal diagnostic sequence or a set of ISO requirements.
5. Complete required testing before return to routine use
After a failure, the sterilizer should remain out of routine use while the required inspection, corrective action, and testing are completed. CDC guidance describes inspection and successful air-removal testing before use; depending on the circumstances, manufacturer instructions and facility procedures may also call for additional testing or other verification before the sterilizer is returned to routine service.[1] Therefore, a subsequent passing Bowie-Dick result should not be presented as universally sufficient by itself.
Document the test results, technical findings, corrective action, and authorization to return the sterilizer to service according to the facility’s process. The applicable decision should be based on the complete documented assessment, not on the indicator result alone.
6. Assess loads processed after the last acceptable test
The scope of any hold, review, or recall should be determined through the facility’s documented nonconformity and recall process, together with applicable requirements and qualified infection-prevention or technical advice. Relevant information may include the time of the failed test, the last acceptable test, cycle records, alarms, chemical-indicator results, load contents, and other evidence concerning process control.
A failed Bowie-Dick test is not, by itself, proof that every earlier load was nonsterile. It is a significant process signal that warrants controlled assessment rather than informal release. The decision about affected loads and their disposition is a facility-process determination, not one that can be made from the Bowie-Dick result alone.
Structuring the investigation
Facilities may improve consistency by using a short, standardized investigation checklist. As a local quality-system tool, it can capture:
- the observed indicator pattern and its interpretation;
- test-system identification, lot information, expiry, and storage status, when applicable;
- sterilizer, chamber, cycle, operator, and timestamp;
- the cycle printout or electronic record;
- recent maintenance, relocation, repair, or utility interruption;
- potentially affected loads and their release status; and
- corrective action, retesting, authorization to return to service, and any follow-up review required by facility procedure.
Trend review can also be included when appropriate to the facility’s quality process. A single failure may require immediate action, while intermittent failures across shifts, cycles, or sterilizers may indicate a recurring equipment, utility, training, or supply issue. These checklist and trend-review practices are recommendations for organizing a local investigation, not direct requirements stated in the cited sources.
Important limitations
Interpretation must remain specific to the test system, sterilizer design, cycle, and facility procedure. A test intended for one defined test load should not automatically be treated as interchangeable with another system. ISO 11140-4:2020 addresses the indicator system together with its specified test load.[2]
Local regulations, accreditation requirements, sterilizer instructions, test-system instructions, and the facility’s quality system may impose additional controls. This article does not establish a release policy or replace manufacturer instructions, qualified technical advice, or facility infection-prevention processes.
Practical takeaway
Use the Bowie-Dick test as an air-removal and steam-penetration challenge for the defined test conditions. A passing result supports the conclusion that the sterilizer passed that particular challenge; it does not provide a complete assessment of every sterilized load. A failed result is a reason to stop routine use, preserve the available evidence, verify test validity, escalate the issue, complete the inspection and testing required by the applicable process, and keep the sterilizer out of routine use until return-to-service criteria are documented.[1]
Educational disclaimer: This article is for professional education and does not replace applicable regulations, standards, sterilizer instructions for use, test-system instructions, maintenance procedures, facility quality processes, or qualified technical and infection-prevention advice.
References
- Steam Sterilization. Centers for Disease Control and Prevention. 2024.
- ISO 11140-4:2020 — Part 4: Type 2 indicator systems as an alternative to the Bowie and Dick-type test. International Organization for Standardization. 2020.
- Premarket Notification [510(k)] Submissions for Chemical Indicators: Guidance for Industry and FDA Staff. U.S. Food and Drug Administration. 2007.
Educational overview. Follow the applicable product instructions, validated procedures, local regulations and your facility’s approved policies.